www.rkiinstruments.com35-3001-01HHydrogen Sample-Draw DetectorOperator’s ManualPart Number: 71-0320RKRevision: P2Released: 9/17/14
6 • 35-3001-01H Hydrogen Sample-Draw DetectorInstallationThis section describes procedures to mount the sample-draw detector in the monitoring enviro
35-3001-01H Hydrogen Sample-Draw Detector • 72. Close and latch the housing door.NOTE: The sample-draw detector is shipped with the mounting feet “t
8 • 35-3001-01H Hydrogen Sample-Draw Detectorring, if it came out, in the orientation shown in Figure 4. 4. Screw the inlet fitting tube nut back ont
35-3001-01H Hydrogen Sample-Draw Detector • 9Installing the Exhaust Line1. Loosen the nut on the exhaust fitting until 3 threads are visible.2. Push
10 • 35-3001-01H Hydrogen Sample-Draw Detector8. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown i
35-3001-01H Hydrogen Sample-Draw Detector • 11Start UpThis section describes procedures to start up the sample-draw detector and place the sample-dr
12 • 35-3001-01H Hydrogen Sample-Draw DetectorMaintenanceThis section describes maintenance procedures. It includes preventive maintenance procedures
35-3001-01H Hydrogen Sample-Draw Detector • 135. Store the calibration kit in a safe place.Quarterly CalibrationCalibrate the sample-draw detector a
14 • 35-3001-01H Hydrogen Sample-Draw DetectorRecommended action1. Verify that the calibration cylinder contains an adequate supply of a fresh test s
35-3001-01H Hydrogen Sample-Draw Detector • 15particle filter is pointing toward the inlet fitting.4. Reinstall the tubing routed to the sampling ar
35-3001-01H Hydrogen Sample-Draw DetectorWARNINGRead and understand this instruction manual before operating detector. Improper use of the detector c
16 • 35-3001-01H Hydrogen Sample-Draw DetectorCalibrationThis section describes how to calibrate the hydrogen sample-draw detector. It includes proce
35-3001-01H Hydrogen Sample-Draw Detector • 17Parts ListTable 2 lists replacement parts and accessories for the sample-draw hydrogen detector.Table
35-3001-01H Hydrogen Sample-Draw DetectorProduct WarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in mat
35-3001-01H Hydrogen Sample-Draw DetectorTable of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
35-3001-01H Hydrogen Sample-Draw Detector • 1OverviewThis operator’s manual describes the 35-3001-01H hydrogen sample-draw detector. This manual als
2 • 35-3001-01H Hydrogen Sample-Draw DetectorDescriptionThis section describes the components of the hydrogen sample-draw detector. Figure 1: Hydro
35-3001-01H Hydrogen Sample-Draw Detector • 3installation where general purpose equipment is in use.The housing door is hinged on the left side and
4 • 35-3001-01H Hydrogen Sample-Draw DetectorDetector/Amp Terminal StripThe detector/amp terminal strip is the upper twelve-point terminal strip in t
35-3001-01H Hydrogen Sample-Draw Detector • 5Status LEDsTwo status LEDs are above the flowmeter. They are also visible through the window in the hou