164 Using Maintenance Mode Gas Tracer Operator’s Manual
Updating the Bump Test Interval
The BP INTVL setting defines the amount of time between bump
tests. The minimum setting is 0 days and the maximum setting is 30
days. The factory setting is 30 days.
NOTE: If the BUMP DSP menu item is set to OFF, this menu item
does not appear.
1. Use the AIR or (SHIFT) buttons to display the BP INTVL
menu item. Press and release the POWER ENTER button. The
screen indicates the current setting.
2. Use the AIR or (SHIFT) button to display the desired setting.
Tip: Press and hold the AIR or (SHIFT) button to rapidly scroll
through the settings.
3. Press and release the POWER ENTER button to save the setting
and return to the main menu.
30
MAINTENANCE
BP INTVL
Zobrazit stránku 171