Rki-instruments GX-2003 Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Senzory Rki-instruments GX-2003. RKI Instruments GX-2003 User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - Operator’s Manual

www.rkiinstruments.com GX-2003Data Logging SoftwareOperator’s Manual Part Number: 71-0096RKRevision: AReleased: 4/12/10

Strany 2 - Warranty

10 • IrDA Downloading CableIrDA Downloading CableThe GX-2003 communicates with a computer via an on-board infrared data port that complies with IrDA

Strany 3 - Table of Contents

IrDA Downloading Cable • 11Logging Software on a Windows® computer, it is necessary to make a special setting in the Wireless Link Configuration windo

Strany 4 - 4 • Table of Contents

12 • IrDA Downloading Cable 6. Click OK. 7. Close the Control Panel window.Using a USB or Serial Port With an IrDA Cable InstalledWhen an IrDA cable

Strany 5 - Introduction

Launching the Software • 13Launching the Software 1. Click Start on the Windows® Icon Tray, then select Programs, then select the GX-2003 folder, the

Strany 6 - System Requirements

14 • Control ButtonsControl ButtonsThis section provides a description of the control buttons. Instructions for using the software are given in other

Strany 7 - Logging Software

Control Buttons • 15Download ButtonIf you are in another program window, clicking the Download button opens the Download window (see Figure 4). The D

Strany 8

16 • Control Buttonsand the fields in the Instrument Information window will become blank.Data ButtonClicking the Data button opens the Data window.In

Strany 9

Control Buttons • 17The following types of data files are saved in the Data window:• Calibration History FilesA calibration history file is saved for e

Strany 10 - IrDA Downloading Cable

18 • Control ButtonsLast Calibration ButtonClicking the Last Calibration button opens the Last Calibration window.The Last Calibration window stores

Strany 11 - IrDA Downloading Cable • 11

Control Buttons • 19• Need Calibration — selecting this option shows calibration information only for the GX-2003 units due for calibration.NOTE: The

Strany 12 - Installed

2 • Warranty Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and perfo

Strany 13 - Launching the Software

20 • Control ButtonsNOTE: The GX-2003 units that are due for calibration will have calibration dates highlighted in red for the channels that are due

Strany 14 - Control Buttons

Control Buttons • 21The fields also include the following: • Gas — Lists the target gas for which the Before, After, and A.Cal readings are displayed.

Strany 15 - Instrument Information Button

22 • Control ButtonsNOTE: If a unit is calibrated using the Single Calibration menu in the GX-2003 (see the GX-2003 instruction manual) it is possibl

Strany 16 - Data Button

Control Buttons • 23Set ButtonClicking the Set button opens the Set window.In the Set window, you can perform the following functions:• Alter the app

Strany 17 - Control Buttons • 17

24 • Accessing a Common DatabaseAccessing a Common DatabaseThe GX-2003 Data Logging Software stores downloaded data in a database file. This database

Strany 18 - Last Calibration Button

Accessing a Common Database • 25very last line. This line is the Data Location path. An example of the path created by the software is shown in Figur

Strany 19 - Control Buttons • 19

26 • Accessing a Common Database 4. Save the file and reopen the GX-2003 Data Logging Software. 5. When you launch the program again, a database will

Strany 20 - 20 • Control Buttons

Downloading Data from the GX-2003 • 27Downloading Data from the GX-2003You have the option of downloading data manually or automatically. If you want

Strany 21 - Control Buttons • 21

28 • Downloading Data from the GX-2003infrared port on the side of the GX-2003 with the infrared port on the cable. 3. Press and hold the POWER/ENTER

Strany 22 - 22 • Control Buttons

Downloading Data from the GX-2003 • 29•To download all data and instrument information from the instrument, click Complete Download.•To download inst

Strany 23 - Set Button

Table of Contents • 3Table of ContentsWarranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 24 - Database

30 • Downloading Data from the GX-2003 7. If the unit has been downloaded manually (Automatic Download not selected) using the Complete Download or I

Strany 25

Downloading Data from the GX-2003 • 31 8. You can view, print, export, or delete downloaded data by entering the Data or Last Calibration windows. Th

Strany 26

32 • Viewing, Printing, Exporting, and Deleting DataViewing, Printing, Exporting, and Deleting DataThe GX-2003 logs four types of data files: calibrat

Strany 27 - Downloading Data from the

Viewing, Printing, Exporting, and Deleting Data • 33 The Data window is divided into four frames. The upper left frame is the Data frame and displays

Strany 28

34 • Viewing, Printing, Exporting, and Deleting Dataselected in the Data frame. The lower right frame shows the summary information for a data file th

Strany 29

Viewing, Printing, Exporting, and Deleting Data • 35 2. If necessary, double-click the GX-2003 icon in the top of the Data window’s upper left frame

Strany 30

36 • Viewing, Printing, Exporting, and Deleting Databe shown in the bottom right frame along with the total number of calibrations saved if it is mor

Strany 31

Viewing, Printing, Exporting, and Deleting Data • 37After specifying the, file name, file location, and file type click the Save button to save the file

Strany 32 - Data Window

38 • Viewing, Printing, Exporting, and Deleting DataThe data may be viewed either in table format or graph format if at least 5 scheduled data points

Strany 33

Viewing, Printing, Exporting, and Deleting Data • 39 4. Click the expanded view symbol (+) of or double-click the dated folder whose contents you wan

Strany 34 - Calibration History

4 • Table of ContentsDeleting Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53Changing the Password . . .

Strany 35

40 • Viewing, Printing, Exporting, and Deleting Dataaverage gas readings over the user defined interval. So if the data logging session started at 4:1

Strany 36

Viewing, Printing, Exporting, and Deleting Data • 41• If you click the Summary button, the data window will split into two frames, one above the othe

Strany 37 - Interval Trend Data

42 • Viewing, Printing, Exporting, and Deleting DataWhen viewing interval trend data in graph format, you have several options: •You can choose which

Strany 38

Viewing, Printing, Exporting, and Deleting Data • 43• If you click the Summary button, the data window will split into two frames one above the other

Strany 39

44 • Viewing, Printing, Exporting, and Deleting Data 12.To save the table or graph to the Clipboard, click the To clipboard button and the table or g

Strany 40

Viewing, Printing, Exporting, and Deleting Data • 45To view and perform desired operations with the alarm trend files: 1. With the software already l

Strany 41

46 • Viewing, Printing, Exporting, and Deleting Dataappear in the Name column. A prefix of “al” indicates an alarm trend data file. 6. Click one of the

Strany 42 - Zoom level

Viewing, Printing, Exporting, and Deleting Data • 47• The gas readings at the time of the alarm event around which the logged data are centered are h

Strany 43

48 • Viewing, Printing, Exporting, and Deleting DataWhen viewing alarm trend data in graph format, you have all the same options as when viewing inte

Strany 44 - Alarm Trend Data

Viewing, Printing, Exporting, and Deleting Data • 49to specify the filename, file location, and file type.For Graph view the export file type is Windows

Strany 45

Introduction • 5IntroductionUsing an advanced detection system consisting of up to five gas sensors, the GX-2003 Personal Gas Monitor detects the pres

Strany 46

50 • Viewing, Printing, Exporting, and Deleting DataEvent DataThe GX-2003 not only saves trend files which include logged data at scheduled times, but

Strany 47

Viewing, Printing, Exporting, and Deleting Data • 51 2. If necessary, double-click the GX-2003 icon in the top of the Data window’s upper left frame

Strany 48 - Zoom factor

52 • Viewing, Printing, Exporting, and Deleting Data 7. If you want to print the data, click the Print button. A Print dialog box will appear for you

Strany 49

Viewing, Printing, Exporting, and Deleting Data • 53After specifying the, file name, file location, and file type click the Save button to save the file

Strany 50 - Event Data

54 • Viewing, Printing, Exporting, and Deleting DataTo delete calibration data in the Last Calibration window: 1. With the software launched, click

Strany 51

Viewing, Printing, Exporting, and Deleting Data • 55 6. Type the new password again, then click the Confirm New Password button. 7. Click OK when the

Strany 52

Calibrating A GX-2003 With the Data Logging Software • 56Calibrating A GX-2003 With the Data Logging SoftwareA GX-2003 can be calibrated using the GX

Strany 53 - Deleting Data

Calibrating A GX-2003 With the Data Logging Software • 57To calibrate a GX-2003, perform the following steps: 1. Launch the GX-2003 Data Logging Sof

Strany 54 - Changing the Password

Calibrating A GX-2003 With the Data Logging Software • 58 6. Use the selection boxes under the Span column to select the channels you want to calibra

Strany 55 - Password button

Calibrating A GX-2003 With the Data Logging Software • 59 8. Verify that the value(s) in the Auto Cal. column match the gas concentration(s) in the c

Strany 56 - Calibrating A GX-2003 With

6 • System RequirementsCAUTION: The GX-2003 detects oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide, and hydrogen

Strany 57

Changing Data Logging Parameters • 60Changing Data Logging ParametersTo make changes to parameters stored in the GX-2003, use the Set window of the

Strany 58

Changing Data Logging Parameters • 61 5. To change the serial number, station ID, or user ID stored in the GX-2003, click the field you wish to change

Strany 59

Changing the Appearance of the Software Screens • 62Changing the Appearance of the Software ScreensTo change how information is displayed in the sof

Strany 60 - Parameters

Changing the Appearance of the Software Screens • 63 4. Select the colors used to graph the various target gases in the interval and alarm trend files

Strany 61

Spare Parts List • 64Spare Parts ListTable 1: Spare Parts ListPart Number Description47-5026RK Cable, serial/IrDA adapter49-5027RK Cable, USB/IrDA ad

Strany 62 - Changing the Appearance

Installing the GX-2003 Data Logging Software • 7Installing the GX-2003 Data Logging Software 1. Launch Windows®. 2. Exit from all applications and op

Strany 63

8 • Installing the GX-2003 Data Logging Software 5. Review the Confirmation section to make sure that your Windows® operating system has the minimum r

Strany 64 - Spare Parts List

Installing the GX-2003 Data Logging Software • 9 8. The installation process will begin. If a File Download window appears asking if you want to open

Komentáře k této Příručce

Žádné komentáře