Gas Detector Head
GD-70D Series
Operating Manual
Request for the Customers
Read and understand this operating manual before using the
detector.
You must operate the detector in accordance with the operating
manual.
Regardless of warranty period, we shall not make any
compensation for accidents and damage caused by using this
product. The compensation shall be made only under the warranty
policy of products or parts replacement.
Because this is a safety device, a regular maintenance for every
six months and daily maintenance must be performed.
If you find abnormalities in the detector, please notify them to our
local representative immediately.
RKI Instruments, Inc. 33248 Central Ave. Union City, CA 94587
Phone: 800-754-5165 Fax: 510-441-5650
Zobrazit stránku 0
1 2 3 4 5 6 ... 80 81

Shrnutí obsahu

Strany 1 - GD-70D Series

Gas Detector Head GD-70D Series Operating Manual Request for the Customers • Read and understand this operating manual before using the detector

Strany 2 - Preface

1 Overview 1-4. Names and functions for each part GD-70D - 10 - <Front Panel and Character LCD> ① MODE key U

Strany 3 - <Contents>

1 Overview 1-5. Units description - 11 - GD-70D 1-5. Units description The detector consists of the following units. &l

Strany 4 - Important Notices on Safety

1 Overview 1-5. Units description GD-70D - 12 - CAUTION • The sensor unit must be handled carefully to ensure quality as safety device. W

Strany 5 - CAUTION

1 Overview 1-5. Units description - 13 - GD-70D Semiconductor type (SGU) • Before using the sensor unit, it needs to be warmed up for a sp

Strany 6 - Overview

1 Overview 1-6. Block diagram GD-70D - 14 - <Wall-mounted Unit> 1-6. Block diagram <Electric Diagram>

Strany 7 - 1-3. Product specifications

1 Overview 1-6. Block diagram - 15 - GD-70D <Tubing Diagram>

Strany 8 - GD-70D - 8

2 Installation 2-1. Requirements GD-70D - 16 - CAUTION • This is a precision instrument. Because the detector may not provide the specifi

Strany 9 - <Appearance>

2 Installation 2-1. Requirements - 17 - GD-70D Do not install the detector in a place exposed to direct sunlight or sudden changes in the temperatu

Strany 10

2 Installation 2-1. Requirements GD-70D - 18 - Introducing protective measures against lightning If cables are installed outside the factory/pl

Strany 11 - 1-5. Units description

2 Installation 2-2. Installation of detector - 19 - GD-70D GD-70D SK1 SK2 Power supply Pow

Strany 12 - CAUTION

Preface GD-70D - 2 - Preface Thank you for choosing our gas detector head GD-70D Series. Please verify that the model number of the product y

Strany 13 - - 13 - GD-70D

2 Installation 2-2. Installation of detector GD-70D - 20 - <Install Dimension and Maintenance Space>

Strany 14 - 1-6. Block diagram

2 Installation 2-2. Installation of detector - 21 - GD-70D <Installation of Wall-mounted Unit> Attach the wall-mounted un

Strany 15 - - 15 - GD-70D

2 Installation 2-2. Installation of detector GD-70D - 22 - <Detaching and Attaching Main Unit> Attaching Main Unit At the position of 10

Strany 16 - Installation

2 Installation 2-3. How to wire - 23 - GD-70D 2-3. How to wire NOTE To use the pyrolyzer unit (PLU-70), also refer to the individual operating ma

Strany 17

2 Installation 2-3. How to wire GD-70D - 24 - Connection conditions • Cable: 0.08 - 2.5mm2 • Bare wire length: 8 - 9 mm • Connecting tool: D

Strany 18

2 Installation 2-3. How to wire - 25 - GD-70D <How to Connect to Terminal Plate> When cables are connected to the connectors, use the dedica

Strany 19

2 Installation 2-3. How to wire GD-70D - 26 - Grounding Use the grounding terminal to connect the detector to your grounding circuit.

Strany 20 - (Unit: mm)

2 Installation 2-3. How to wire - 27 - GD-70D Wiring Example Connecting to the indicator (3-wire type - 4 - 20 mA specification) Connec

Strany 21 - - 21 - GD-70D

2 Installation 2-3. How to wire GD-70D - 28 - <How to tube> NOTE To use the pyrolyzer unit (PLU-70), also refer to the individual operat

Strany 22

2 Installation 2-4. Relocate - 29 - GD-70D 2-4. Relocate When the detector is relocated, select a new place in accordance with "Precautions f

Strany 23 - 2-3. How to wire

- 3 - GD-70 <Contents> Preface ...

Strany 24 - Dedicated

3 How to Operate 3-1. Preparation for start-up GD-70D - 30 - 3 How to Operate 3-1. Preparation for start-up Before connecting a power

Strany 25 - Position the

3 How to Operate 3-2. How to start the detector - 31 - GD-70D <Start-up Procedures (approximately 25 seconds for system check of the detector an

Strany 26 - WARNING

3 How to Operate 3-3. How to exit GD-70D - 32 - 3-3. How to exit To turn off the detector, open the switch cover on the bottom of the main uni

Strany 27 - - 27 - GD-70D

3 How to Operate 3-5. Modes - 33 - GD-70D <Detection mode> Flow Rate Indicator Because the suction flow rate of the detector is auto

Strany 28 - <How to tube>

3 How to Operate 3-5. Modes GD-70D - 34 - Mode Item LCD Display Details Regular Maintenance Mode Switching 1-8 M MODE Switch to the regular

Strany 29 - 2-5. Disposal

4 Detection Mode 4-1. Gas alarm activation - 35 - GD-70 4 Detection Mode 4-1. Gas alarm activation Gas alarm: Activated when the concen

Strany 30 - How to Operate

4 Detection Mode 4-1. Gas alarm activation GD-70D - 36 - <Contact Activation> The contact is activated when the gas concentration reache

Strany 31 - WARM UP

4 Detection Mode 4-2. Fault alarm activation - 37 - GD-70D <Response to Gas Alarm> In case of responding to a leaked gas When a gas alarm is

Strany 32 - GD-70D - 32

4 Detection Mode 4-3. External output operation GD-70D - 38 - 4-3. External output operation Specifications 4 - 20 mA Power-line communicati

Strany 33 - 3-5. Modes

4 Detection Mode 4-3. External output operation - 39 - GD-70D Example of Gas Concentration and External Output 4 - 20 mA Specification (Maintenance

Strany 34 - GD-70D - 34

Important Notices on Safety GD-70 - 4 - Important Notices on Safety <Definition of DANGER, WARNING, CAUTION, and NOTE> DANGER Indicates

Strany 35 - Detection Mode

4 Detection Mode 4-4. Other functions GD-70D - 40 - 4-4. Other functions <Suppression Function> Some types of sensor used with the dete

Strany 36 - <Contact Activation>

4 Detection Mode 4-4. Other functions - 41 - GD-70D <Zero Follower Function> Some types of sensor used with the detector might have sensitiv

Strany 37 - 4-2. Fault alarm activation

4 Detection Mode 4-4. Other functions GD-70D - 42 - <Calibration History/Alarm Trend History/Event History Functions> The detec

Strany 38 - GD-70D - 38

5 Alarm Test Mode - 43 - GD-70 5 Alarm Test Mode This is used when dummy signals the same as the signals of the gas concentration are gene

Strany 39

6 User Mode GD-70 - 44 - 6 User Mode User Mode PW A1 A2 F LCD Detection Mode Press the MODE key for three seconds. ● ○ ○

Strany 40 - 4-4. Other functions

6 User Mode - 45 - GD-70D 1-5.70D VER Show the program version of the main unit. ● ○ ○ ○1- 5 70D VER MAINTENANCE → SET01234 56ABMAINTENANCE ▲

Strany 41 - Function>

6 User Mode GD-70D - 46 - <Zero Adjustment "1-1"> This is used to perform the zero adjustment. Before starting the zero adjus

Strany 42 - GD-70D - 42

6 User Mode - 47 - GD-70D <Setpoint Indicator "1-2"> This is used to check important setpoints. Setpoint Display PW A1 A2 F L

Strany 43 - PW A1 A2 F

7 Regular Maintenance Mode GD-70 - 48 - 7 Regular Maintenance Mode Mode Item LCD Display Details Gas Introduction Display 2-0

Strany 44 - User Mode

7 Regular Maintenance Mode - 49 - GD-70D Mode Item LCD Display Details Environmental Setting 2 ⇒P55 2-10 SETTING2 Functions setting • Address

Strany 45

Important Notices on Safety - 5 - GD-70D <Precautions> CAUTION Do not use a transceiver (walkie-talkie) near the detector. Radio wave from a

Strany 46 - PW AL1 AL2 F LCD

7 Regular Maintenance Mode GD-70D - 50 - Regular Maintenance Mode PW A1 A2 F LCD User Mode In "1-8.M MODE", press the SET key. ●

Strany 47 - PW A1 A2 F LCD

7 Regular Maintenance Mode - 51 - GD-70D ▲↓ ↑▼ ● ○ ○ ○2- 5 DEF FLOWMAINTENANCE ⇔ SET 1000 DEF FLOWMAINTENANCE 2-5.DEF FLOW Set the flow r

Strany 48 - Regular Maintenance Mode

7 Regular Maintenance Mode GD-70D - 52 - ▲↓ ↑▼ ● ○ ○ ○2-12 FAULT MAINTENANCE 2-12.FAULT This is used (by the manufacturer) to inves

Strany 49

7 Regular Maintenance Mode - 53 - GD-70D <Environmental Setting 1 "2-9"> In the environmental setting 1, specify the operation se

Strany 50

7 Regular Maintenance Mode GD-70D - 54 - <Alarm Value Setting "2-9" - "SET 1"> Alarm Value Setting PW A1 A2 F LCD

Strany 51

7 Regular Maintenance Mode - 55 - GD-70D <Environmental Setting 2 "2-10"> In the environmental setting 2, specify the settings of

Strany 52

7 Regular Maintenance Mode GD-70D - 56 - ▲↓ ↑▼ ● ○ ○ ○SET 5 TEST4-20 MAINTENANCE ⇔ SET ON TEST4-20 MAINTENANCE ▲↓↑▼ OFF TES

Strany 53 - Environmental Setting 1

7 Regular Maintenance Mode - 57 - GD-70D ▲↓ ↑▼ ● ○ ○ ○SET 10 FLT PTRN MAINTENANCE SET 10. FLT PTRN This is a setting screen of the faul

Strany 54 - PW A1 A2 F LCD

7 Regular Maintenance Mode GD-70D - 58 - ▲↓ ↑▼ ● ○ ○ ○SET 15 MNT OUT MAINTENANCE ⇔ SET 2.5mA MNT OUT MAINTENANCE ▲↓↑▼ 4.0mA

Strany 55 - Environmental Setting 2

7 Regular Maintenance Mode - 59 - GD-70D ● ○ ○ ○SET 19 PUMP CK MAINTENANCE ⇔ SET ON PUMP CK MAINTENANCE ▲↓↑▼ OFF PUMP CK MAINTENANCE S

Strany 56

1 Overview 1-1. Product components GD-70 - 6 - 1 Overview 1-1. Product components <Gas Detector (GD-70D Series)> <Standard

Strany 57

7 Regular Maintenance Mode GD-70D - 60 - <Date/Time Setting "2-10" - "SET 1"> Date/Time Setting PW A1 A2 F LCD S

Strany 58

7 Regular Maintenance Mode - 61 - GD-70D <Energized/De-Energized Contact Setting "2-10" - "SET 6"> Energized/De-Energiz

Strany 59 - PUMP CK

8 Maintenance 8-1. Maintenance intervals and items GD-70 - 62 - 8 Maintenance This is a critical instrument for disaster-proof and safet

Strany 60 - PW A1 A2 F LCD

8 Maintenance 8-1. Maintenance intervals and items - 63 - GD-70D <About Maintenance Services> • We provide services on regular maintenance

Strany 61 - PW A1 A2 F LCD

8 Maintenance 8-2. Gas calibration method GD-70D - 64 - 8-2. Gas calibration method Perform a gas calibration in each mode (zero adjustment mo

Strany 62 - Maintenance

8 Maintenance 8-2. Gas calibration method - 65 - GD-70D Zero Adjustment PW A1 A2 F LCD 2-1.ZERO Press the SET key. ● ○ ○ ○2- 1 ZERO MAINTENANCE

Strany 63

8 Maintenance 8-2. Gas calibration method GD-70D - 66 - <Span Adjustment "2-2"> This is used to perform the span adjustment. F

Strany 64 - Calibration

8 Maintenance 8-3. Other adjustments/Cleaning method - 67 - GD-70D 8-3. Other adjustments/Cleaning method <Flow Rate Manual Adjustment "2-

Strany 65

8 Maintenance 8-4. How to replace parts GD-70D - 68 - <Cleaning of Detector> Clean the detector if it becomes extremely dirty. The detec

Strany 66

8 Maintenance 8-4. How to replace parts - 69 - GD-70D <Replacement of Regular Replacement Parts> List of recommended regular replacement par

Strany 67

1 Overview 1-3. Product specifications - 7 - GD-70D 1-3. Product specifications <Common Specifications> Concentration Display Character LCD

Strany 68 - 8-4. How to replace parts

8 Maintenance 8-5. Procedures to store the detector or leave it for a long time GD-70D - 70 - 8-5. Procedures to store the detector or leave it

Strany 69 - Replacement of Flow Sensor

9 Troubleshooting - 71 - GD-70 9 Troubleshooting The troubleshooting does not explain the causes of all the malfunctions which occur on th

Strany 70

9 Troubleshooting GD-70D - 72 - Symptom/Display FAULT Causes Actions Zero drift caused by environmental changes or aging degradation is out

Strany 71 - Troubleshooting

9 Troubleshooting - 73 - GD-70D Symptom/Display FAULT Causes Actions The rated voltage is not supplied to the detector. Check the power supply, an

Strany 72 - - 72

Definition of Terms GD-70 - 74 - Definition of Terms External Dust Filter When the detector is used in a dusty environment, it is recommended t

Strany 73

Detection principle - 75 - GD-70 Detection principle <Electrochemical Type> The electric potential between the working electrode and refe

Strany 74 - Definition of Terms

Detection principle GD-70D - 76 - <Pyrolysis-Particle Type> When the gas to be detected is heated to several hundred degrees, particula

Strany 75 - Detection principle

Detection principle - 77 - GD-70D The detector has the pyrolysis-particle type sensor, which is a radioisotope-equipped device. It is examined in

Strany 76 - GD-70D - 76

Detection principle GD-70D - 78 - Safety of the radioisotope 241Am (37 KBq) used in the pyrolysis-particle type sensor The pyrolysis-particle

Strany 77 - - 77 - GD-70D

Detection principle - 79 - GD-70D <Semiconductor Type> Metal dioxide can measure gas concentration based on changes in the electric conduct

Strany 78 - GD-70D - 78

1 Overview 1-3. Product specifications GD-70D - 8 - <Specifications for Each Principle> ESU ESU+PLU *1 SSU+PLU *1 SGU OSU *2 Detection

Strany 79 - <Semiconductor Type>

Detection principle GD-70D - 80 - <Galvanic Cell Type> By immersing precious metal and lead in electrolyte and connecting them with a l

Strany 80 - <Galvanic Cell Type>

- 81 - GD-70 Warranty Policy 1. We shall provide free repair services for one year after your purchase of the product in case of malfunctions,

Strany 81 - Warranty Policy

1 Overview 1-4. Names and functions for each part - 9 - GD-70D 1-4. Names and functions for each part <Appearance>