GX-2012 Operator’s Manual Using Maintenance Mode 125
set to OFF, the Clear Data Logger Screen does not appear in Display Mode
and any logged data cannot be cleared from Display Mode.
1. Use the AIRS or (SHIFT)T buttons to display the DLOG DSP menu
item.
2. Press and release the POWER ENTER button. The screen will indicate
the current setting.
3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Turning the Bump Test Function On or Off
The BUMP DSP menu item is used to turn the bump test function on or off.
If set to OFF, the BUMP menu item in Calibration Mode and Maintenance
Mode does not appear. In addition, the BUMP--SET, BP INTVL, BP
RMNDR, and BP EXPRD menu items will not display in Maintenance
Mode. The factory setting is OFF.
1. Use the AIRS or (SHIFT)T buttons to display the BUMP DSP menu
item.
MAINTENANCE
DLOG DSP
On
MAINTENANCE
DLOG DSP
MAINTENANCE
BUMP DSP