22 Start Up, Normal Mode GX-2012 Operator’s Manual
If the unit is due for a bump test and BP EXPRD is set to NOT USE,
then the following screen displays.
The GX-2012 cannot be used until a bump test has been performed
by selecting BUMP in the Maintenance Mode menu. See
“Performing a Bump Test” on page 111 for bump testing instructions.
If a bump test is not due or BP EXPRD is set to NO EFFECT, then
the following screen appears for a few seconds indicating when the
next bump test is due.
NOTE: If both CL RMNDR and BP RMNDR are turned off, a WARM UP
screen will display before the warm up sequence continues.
FAIL
B--LIMIT
2011
9
30
10d
NEXTBUMP
Zobrazit stránku 29