35-3000RK-LEL/O Sample-Draw Detector 13
6. If the fail condition continues, contact RKI Instruments, Inc. for further instruction.
Slow or No Response/Difficult or Unable to Calibrate
Symptoms
The sensors respond slowly or do not respond during the monthly response test.
Unable to accurately set the zero or response reading during the calibration
procedure.
The sensors require frequent calibration.
Probable causes
The calibration cylinder is low, out-dated, or defective.
If a demand flow regulator calibration kit is used, the demand flow regulator is not
functioning properly.
The sample-draw detectors flow rate is too low because of an obstructed sample line,
failed pump, etc.
The sample-draw detector is malfunctioning.
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If a demand flow regulator calibration kit is used, use a different demand flow
regulator to determine if the original one is functioning properly.
3. If necessary, set the correct flow rate with the bypass valve or flow adjust
potentiometer.
4. If you cannot set the correct flow rate, check the sample line for obstructions or kinks.
5. If the calibration/response difficulties continue, replace the sensors as described later
in this section.
6. If the calibration/response difficulties continue, contact RKI Instruments, Inc. for
further instruction.
Replacing Components of the Sample-Draw Detector
This section includes procedures to replace the sensors, hydrophobic filter, and ferrules.
Replacing the Combustible Gas Sensor
1. Turn off the controller
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Unscrew and remove the two screws that secure the sensor retaining plate, then lift
the plate, connector, and sensor out of the housing.
5. Unplug the connector from the sensor.
6. Verify that you are using the correct replacement sensor (NC-6240 is printed on the
sensor), then plug the sensor into the connector.
7. Place the sensor in the combustible gas sensor cavity, then position the retaining plate
on the two standoffs.
8. Secure the retaining plate to the standoffs with the two screws you removed in step 4.
9. Close and latch the housing door.
10. Turn on power to the controller.